Maintaining the FID 11
Maintaining Your GC 223
17
Lower the collector assembly into the housing.
18 Insert the three screws and tighten (to 18 inch- pounds).
19 Connect the ignitor extension cable.
20 Verify assembly:
a Check the FID leakage current. (See “To Check the FID
Leakage Current” on page 224.)
b Bakeout the detector. (See “To Bakeout the FID” on
page 229.)
c Check the FID baseline. (See “To Check the FID
Baseline” on page 225.)
21 Reset the EMF counters. See To Reset an EMF Counter in
the Operation Manual.