Maintaining the uECD 13
Maintaining Your GC 253
If the id is less the 200 um, mark the column with a
septum 70 ± 1 mm from the end. Insert column and nut
into the adapter with the septum at the rear of the
column nut, and tighten the column nut with one
5/16- inch wrench on the adapter and another 1/4- inch
wrench on the column nut.
9 After heating the detector, retighten the 9/16- inch
makeup adapter nut and 1/4- inch column nut.
70 ± 1 mm
Septum