26 Maintaining Your GC
3 Maintaining the GC
9
While continuing to measure the flow rate from the tube,
adjust the pressure regulator on the calibration gas cylinder
until the flow rate is approximately 30 mL/min. Remove the
flowmeter and close the oven door.
10 Allow the calibration cycle to complete (approximately
5 minutes total).
11 Check for leaks.