278 Maintaining Your GC
14 Maintaining the NPD
24
Start the Adjust offset process. Enter the desired offset in
the Target offset field. The default offset is 20 pA for Blos
beads, and 30 pA for white or black beads. For white and
black beads, an offset of 25 to 30 pA is sufficient for
most applications. The bead life may be shortened at a
higher offset.
25 Reset the bead counter. See To Reset an EMF Counter in
the Operation Manual.