Maintaining the NPD 14
Maintaining Your GC 287
12
Start the NPD bead Auto Adjust process. See the 7890
Series Advanced User Guide for details.
Allow 4 to 24 hours for a new ceramic bead to equilibrate,
or 1 to 2 hours for a new Blos bead.