Maintaining Capillary Columns 4
Maintaining Your GC 35
4
Cap the detector column fitting.
5 Set a minimum velocity of 30 cm/s, or as recommended
by the column manufacturer. Let gas flow through the
column at room temperature for 15 to 30 minutes to
remove air.
6 Program the oven from room temperature to the
maximum temperature limit for the column. Increase the
temperature at a rate of 10 to 15 °C/min. Hold at the
maximum temperature for 30 minutes.
7 Load the GC maintenance method and wait for the GC to
become ready.
8 Attach the column to the detector. For details, select your
specific detector from the following list:
• To Install a Capillary Column in the FID
• To Install a Capillary Column in the NPD
• To Install a Capillary Column in the TCD
• To Install a Capillary Column in the uECD
• To install a Capillary Column in the FPD Plus
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals