Maintaining Capillary Columns 4
Maintaining Your GC 39
• To Install a Capillary Column in the TCD
• To Install a Capillary Column in the uECD
• To install a Capillary Column in the FPD Plus
• To Install a Capillary Column Adapter in the FPD
8 Set the column flow to the normal operating value, or set
the capillary column gas velocity to 30 cm/s.
For Multimode, Split/Splitless, PTV, and VI inlets, select
split mode and set the split vent flow to 200 mL/min.
9 Purge the column with carrier flow for at least
10 minutes before heating the oven.
10 Set the inlet temperature to 300 °C or 25 °C above the
normal operating temperature.
11 Set the column oven 25 °C above the GC method final
oven temperature to bake contaminants out of the inlet,
mostly through the split vent. Do not exceed the column
manufacturer’s maximum temperature limit.
12 Bakeout for 30 minutes.