Maintaining the Split/Splitless Inlet 5
Maintaining Your GC 55
5
Use tweezers to remove the septum or Merlin Microseal
from the retainer nut. (See “To Change the Septum on the
Split/Splitless Inlet” on page 52.)
6 Scrub the residue from the retainer nut and septum
holder with a small piece of rolled- up steel wool and
tweezers. Do not do this over the inlet.
7 Use compressed air or nitrogen to blow away the pieces
of steel wool and septum.
8 Line up the tab on the bottom of the septum assembly
with the slot on the insert assembly and push down to
connect. Slide the locking tab to the left.
9 Firmly press the new septum or Merlin Microseal into the
fitting. (See “To Change the Septum on the Split/Splitless
Inlet” on page 52.)
10 Replace the septum retainer nut or Merlin cap and
finger- tighten. (See “To Change the Septum on the
Split/Splitless Inlet” on page 52.)
11 Restore the analytical method.
12 Reset the EMF counter. See To Reset an EMF Counter in
the Operation Manual.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals