88 Maintaining Your GC
6 Maintaining the Purged Packed Inlet
11
Blow the inside of the inlet dry with compressed air or
nitrogen.
12 Install the adapter, if used. (See “To Install an Adapter on
the Purged Packed Inlet” on page 81.)
13 Install the glass liner and O- ring. (See “To Change the
Glass Liner on the Purged Packed Inlet” on page 84.)
14 Install the top insert weldment and finger- tighten.
15 Install the septum and septum nut. (See “To Change the
Septum on the Purged Packed Inlet” on page 77.)
16 Attach the column. (See “To Install a Capillary Column
with the Purged Packed Inlet” on page 72.)
17 Check for leaks.
18 Restore the analytical method.
19 Set the septum and glass liner counters.