EasyManuals Logo

Agilent Technologies Agilent 7890 Series User Manual

Agilent Technologies Agilent 7890 Series
354 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Maintaining the Purged Packed Inlet 6
Maintaining Your GC 95
5
Tighten the inlet column nut 1/4 turn with a wrench.
6 After conditioning, remove the column from the inlet.
7 Simultaneously insert the column into the inlet and
detector fittings but do not force it. It may be necessary to
start the long end of the column in the inlet at an angle to
clear the oven floor.
8 Withdraw the column 1 to 2 mm from both the inlet and
detector. Finger- tighten both column nuts.
9 Tighten both column nuts 1/4 turn with a wrench.
10 Press [Config], [Column 1] or [Column 2], enter 0.00 in either
Length or Diameter, and identify the inlet and detector to
which the column is connected.
11 Establish a flow of carrier gas and purge as recommended
by the packing manufacturer. Generally:
• 20 to 30 mL/min for 2- mm id glass or 1/8- inch od
metal columns
• 50 to 60 mL/min for 4- mm id glass or 1/4- inch od
metal columns
12 Restore the analytical method.
• For FPD, immediately turn off the flame.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
CAUTION
Overtightening the column nut or forcing it to bottom in either the
inlet or detector may shatter the column.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 7890 Series and is the answer not in the manual?

Agilent Technologies Agilent 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 7890 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals