Issue 2 - 07/00 4-77
Testing in Simulation Mode - Section 5
TN1R6-N
• To receive AOC (outgoing calls), it must be
CorNet-N
emulating a TE.
The quickest way to test AOC is to use the Available services
one-button test—see Testing services from the ISDN Test Suite
later in this section. However, you cannot use this method to
check the number of charging units which have been sent.
Testing AOC-E
To check AOC-E (Advice of Charge displayed at the end of
the call):
-1-
With aurora
Sonata
emulating a TE, make a self
call then disconnect. The call must be long
enough for charging units to be received—the
actual length required depends on the network.
-2-
Press + 6 to display the ISDN call
information screen, and check that the
Charge value is displayed. See Getting
information about the current or last call in
Chapter 3 section 2.
When the charge information is displayed in terms of currency
rather than units, aurora
Sonata
displays CU after the value.
Testing AOC-D
To check AOC-D (Advice of Charge generated during the
call):
-1-
With aurora
Sonata
emulating a TE, make a self
call.
-2-
With the call still connected, press + 6 to
display the ISDN call information screen, and
check that the Charge value is displayed. See
Getting information about the current or last
call in Chapter 3 section 2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies auroraSonata and is the answer not in the manual?

Agilent Technologies auroraSonata Specifications

General IconGeneral
BrandAgilent Technologies
ModelauroraSonata
CategoryTest Equipment
LanguageEnglish

Related product manuals