92 Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
6 GPIB Communication, Connections and Configuration
Configuring the Agilent 82350 GPIB Interface
7 In the Agilent Connection Expert main window select GPIB0 (see
Figure 37).
Figure 37 Agilent 82350 default settings.
8 Click Change Properties.... In the Agilent 82350 PCI GPIB Interface -
GPIB0 dialog window select or enter the following properties:
VISA interface ID: to GPIB0,
GPIB address: to 30,
check System controller (recommended),
SICL Interface ID: to hp82341 (note that this is case sensitive),
Logical unit: to 7 or 9 and
uncheck Auto- discover instruments connected to this interface (see
Figure 38).
Figure 38 Agilent 82350 interface configuration.
NOTE
The appearance of the dialog slightly differs depending on the type of GPIB board installed.