Troubleshooting and Test Functions 3
1100 Series WS MWS Reference Manual 103
8. Sample location error
✔ Check the plate configuration in the user interface
✔ Ensure the right vials and plates are used
✔ Perform an auto-alignment
✔ Check the connector from the needle carrier to the sample transport unit
✔ Check the sample transport connectors on the main board/sample
transport
✔ Check the connector on the bottom of the sample transport unit
✔ Change the needle (G1367-87200) and the seat (G1367-87101)
✔ Change the needle carrier assembly (G1367-60010)
✔ Change the sample transport unit (G1367-60019)
✔ Change the main board (G1367-69500)