274 1100 Series WS MWS Reference Manual
9 Specifications
Performance Specifications
Tabl e 79 Performance Specifications Agilent 1100 Series Well Plate Sampler
Type Specification
GLP features Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications Controller-area network (CAN). RS232C, APG-remote standard,
optional four external contact closures and BCD vial number output
Safety features Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
Injection range 0.1 100 µl in 0.1 µl increments
Up to 1500 µl with multiple draw (hardware modification required)
Precision Typically < 0.5 % RSD of peak areas from 5 100 µl,
Typically < 1 % RSD from 1 5 µl
Sample viscosity range 0.2 5 cp
Sample capacity 2 × well-plates (MTP) + 10 × 2 ml vials
100x2ml in one tray
40 x 2 ml in half tray
Injection cycle time Typically < 30 s using following standard conditions:
Default draw speed: 200 µl/min
Default eject speed: 200 µl/min
Injection volume: 5 µl
Carry-over Typically < 0.01 % using the following conditions:
Column: 125 x 4 mm Hypersil ODS, 5 µm
Mobile phase: Water/Acetonitrile = 80/20
Flow rate: 1 ml/min
Injection volume: 1 µl caffeine (1 mg/ml), 5 µl water to test carryover
Outside wash of needle before injection: 20 sec with water using
flush port

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1377A and is the answer not in the manual?

Agilent Technologies G1377A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1377A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals