30 1100 Series WS MWS Reference Manual
1 Installing the Sampler
Flow Connections to the Sampler
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
valve.
3 Connect the corrugated waste tube to the seat adapter and the solvent
waste from the leak plane.
4 Ensure that the waste tube is positioned inside the leak channel.
5 Drive the tube from the peristaltic flush pump to the solvent bottle in the
solvent cabinet
6 Seat capillary: see recommendations in “Choice of Seat Capillary" on
page 54
Preparation Sampler is installed in the LC system
Parts required Parts form the accessory kits, see Accessory Kits" on page 18
WARNING
When opening capillary or tube fittings, solvents may leak out. Please observe
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1377A and is the answer not in the manual?

Agilent Technologies G1377A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1377A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals