Repairing the Sampler 4
1100 Series WS MWS Reference Manual 151
SLS Board
Removing the SLS board
1 In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
4 Remove the plate tray from the tray base.
5 Remove the transport assembly, (see“Removing the sample transport" on
page 133).
6 Remove the sampling unit, (see “Removing the sampling unit" on page 134).
7 Remove the SLS board cover.
8 Disconnect the ribbon cable from the SLS-board and unscrew the 2 holding
screws.
9 Remove the SLS board from the inside of the module.
Installing the SLS board
1 Install the new SLS board and the SLS board cover.
2 Install the sampling unit, (see “Installing the sampling unit" on page 135).
3 Install the sample transport, (see “Installing the sample transport" on
page 133)
4 Install the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
5 Replace the plate tray in the tray base.
6 Turn ON the sampler, close the front door, the system will reset
automatically.
7 In the user interface perform a “Sample Transport Self Alignment”.
When required When defective.
Tools required 1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required G1367-66505 SLS board