74 1100 Series WS MWS Reference Manual
3 Troubleshooting and Test Functions
Arm Movement Failed or Arm Movement Timeout
The transport assembly was unable to complete a movement in one of the
axes.
The processor defines a certain time window for the successful completion of
a movement in any particular axis. The movement and position of the
transport assembly is monitored by the encoders on the stepper motors. If the
processor does not receive the correct position information from the encoders
within the time window, the error message is generated.
See figure for axes identification.
Arm Movement 0 Failed: X-axis.
Arm Movement 1 Failed: Z-axis.
Arm Movement 2 Failed: Theta (needle carrier rotation).
Probable Causes
• Mechanical obstruction.
• High friction in transport assembly.
• Defective motor assembly.
• Defective sample transport assembly flex board.
• Defective MTP board.
Suggested Actions
✔ Ensure unobstructed movement of the transport assembly.
✔ Exchange the sample transport assembly.
✔ Exchange the MTP board.