74 1100 Series WS MWS Reference Manual
3 Troubleshooting and Test Functions
Arm Movement Failed or Arm Movement Timeout
The transport assembly was unable to complete a movement in one of the
axes.
The processor defines a certain time window for the successful completion of
a movement in any particular axis. The movement and position of the
transport assembly is monitored by the encoders on the stepper motors. If the
processor does not receive the correct position information from the encoders
within the time window, the error message is generated.
See figure for axes identification.
Arm Movement 0 Failed: X-axis.
Arm Movement 1 Failed: Z-axis.
Arm Movement 2 Failed: Theta (needle carrier rotation).
Probable Causes
Mechanical obstruction.
High friction in transport assembly.
Defective motor assembly.
Defective sample transport assembly flex board.
Defective MTP board.
Suggested Actions
Ensure unobstructed movement of the transport assembly.
Exchange the sample transport assembly.
Exchange the MTP board.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1377A and is the answer not in the manual?

Agilent Technologies G1377A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1377A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals