Troubleshooting and Test Functions 3
1100 Series WS MWS Reference Manual 93
Troubleshooting Guide for the Well Plate Sampler G1367A
The objective of this note is to give some guidance on troubleshooting the
Agilent 1100 series well plate sampler.
Additional information will be added when available. The latest version of this
document can be downloaded from the LSBU homepage.
(http://lsbu.marketing.agilent.com/start/start.asp)
Products - Injection system - G1367A - Technical note
Gather information about the problem
• Instrument serial number?
• Instrument firmware revision and user interface revision?
• When did the problem start?
• What was done/changed prior to the start of the problem?
• What errors or error codes are shown in the Chemstation GUI, and in the
current logbook files? (check especially for errors around the start date of
the problem)
• What errors or error codes are shown in the WPS error logbook?
(Chemstation\HPCHEM\instrument\x\temp\lals.txt)