EasyManuals Logo

Agilent Technologies G1377A User Manual

Agilent Technologies G1377A
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
76 1100 Series WS MWS Reference Manual
3 Troubleshooting and Test Functions
Valve to Mainpass Failed
The injection valve failed to switch to the mainpass position.
The switching of the injection valve is monitored by two microswitches on the
valve assembly. The switches detect the successful completion of the valve
movement. If the valve fails to reach the mainpass position, or if the
microswitch does not close, the error message is generated.
Probable Causes
Valve in an intermediate position between the bypass and mainpass
positions.
Defective injection valve.
Defective MTP board.
Suggested Actions
Turn the WPS main power OFF and ON.
Exchange the injection valve.
Exchange the MTP board.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1377A and is the answer not in the manual?

Agilent Technologies G1377A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1377A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals