234 1100 Series WS MWS Reference Manual
7 Theory of Operation
In this way you can store more than one set of parameters using the same 8-bit
configuration switch twice, for example, for both GPIB and RS-232C.
GPIB Default Addresses
If you just want to change the GPIB address and need a detailed procedure,
refer to the Installing Your Agilent ChemStation System handbook.
Default GPIB address is set to the following addresses:
where 0 means that the switch is down and 1 means that the switch is up.
Tabl e 70 Default Addresses for Agilent Series 1100 Modules
Module Address Binary Address
Pump 22 0 0 0 1 0 1 1 0
FLD 23 0 0 0 1 0 1 1 1
VWD 24 0 0 0 1 1 0 0 0
Agilent 8453A 25 0 0 0 1 1 1 0 1
DAD / MWD 26 0 0 0 1 1 0 1 0
Column compartment 27 0 0 0 1 1 0 1 1
Autosampler 28 0 0 0 1 1 1 0 0
Well-plate sampler no address 000000000000
RID 29 0 0 0 1 1 1 0 1