Troubleshooting and Test Functions 3
1100 Series WS MWS Reference Manual 81
Initialization Failed
The sampler failed to complete initialization correctly.
The sampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined routine. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable Causes
Side door not installed correctly.
Sample transport/sampling unit not aligned correctly
Mechanical obstruction.
Defective sampling unit flex board.
Defective transport assembly flex board.
Defective sampling unit motor.
Defective MTP board.
Suggested Actions
Check if the side door is installed correctly.
Check if the magnet is in place in the side door.
Do an auto-alignment
Ensure unobstructed movement of the transport assembly.
Exchange the defective sampling unit motor.
Exchange the sample transport assembly.
Exchange the MTP board.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1377A and is the answer not in the manual?

Agilent Technologies G1377A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1377A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals