Repairing the Sampler 4
1100 Series WS MWS Reference Manual 145
Exhaust Fan
Removing the exhaust fan
1 In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
4 Remove the plate tray from the tray base.
5 Remove the transport assembly, (see“Removing the sample transport" on
page 133).
6 Remove the MTP main board, (see “Removing the MTP main board" on
page 146).
7 Remove the exhaust fan.
Installing the exhaust fan
1 Install the exhaust fan.
2 Install the MTP main board, (see “Installing the MTP main board" on
page 147).
3 Connect the fan connector to the MTP main board.
4 Install the sample transport (see “Installing the sample transport" on
page 133).
5 Install the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
6 Replace the plate tray in the tray base.
7 Turn ON the sampler, close the front door, the system will reset
automatically.
When required When defective.
Tools required 1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required 3160-4097 Fan