Repairing the Sampler 4
1100 Series WS MWS Reference Manual 147
Installing the MTP main board
1 Install the new board. Ensure the ribbon cables are positioned in the slot in
the board.
2 Reconnect the connectors to the board.
3 Replace the connector screws in the back of the module.
4 Replace the M4 screw at the back of the module.
5 Install the sample transport, (see “Installing the sample transport" on
page 133)
6 Install the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
7 Replace the plate tray in the tray base.
8 Turn ON the sampler, close the front door, the system will reset
automatically.
9 Check the firmware revision of the sampler. If the firmware revision is
older than the current firmware revision of the autosampler, update the
firmware, see “Replacing the Autosampler Firmware" on page 155.
10 Perform a “Sample Transport Self Alignment”.
NOTE
If a new sampler main board is installed, update the serial number information of the
sampler in the user interface, see procedure bellow.