Troubleshooting and Test Functions 3
1100 Series WS MWS Reference Manual 95
Errors which may occur during the turn ON and initialization process
Step 1
Symptom “Failure to turn ON”
No activity when power button is pushed on. Power indicator stays off.
Possible causes
Defective main board
Power supply defective
Suggested actions
Turn the WPS off. Disconnect the power supply from the main board, see if
power indicator becomes green when the WPS is turned-on.
If yes, change the main board (G1367-69500)
If no, change the power supply (0905-2528 Rev G or higher)
Step 2
Symptom “Fan failure”
Early during the turn-on/initialization process a main fan or an exhaust fan
error occurs.
Possible causes
Fan not connected to the main board
Fan connector defective
Fan defective
Main board defective