96 1100 Series WS MWS Reference Manual
3 Troubleshooting and Test Functions
Suggested actions
Make sure the fan is connected to the main board correctly
Examine the fan connector for irregularities. Correct if possible
Replace the defective fan (main fan: 3160-1017, exhaust fan: 3160-4097)
Replace the main board (G1367-69500)
Step 3
Symptom 1 “Main board initialization fails”
Status indicator remains off, but the rest of the turn-on/initialization process
is successful.
Possible causes
The heel of the light pipe is not positioned correctly above the LED on the
main board
The main board is defective
Suggested actions
Correctly position the light pipe
Replace the main board (G1367-69500)
Symptom 2
Status indicator remains off, and the rest of the turn-on/initialization process
fails as well.
Suggested action
Refer to the possible causes and actions for step 5 bellow