140 1100 Series WS MWS Reference Manual
4 Repairing the Sampler
Installing the injection valve assembly
1 Feed the cable of the new valve through to the SUD board.
2 Connect the valve cable to the connector labelled “valve assy”.
3 Install and tighten the three screws which fix the injection valve assembly.
4 Install the sampling unit, (see “Installing the sampling unit" on page 135).
5 Install the sample transport (see “Installing the sample transport" on
page 133).
6 Install the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
7 Replace the plate tray in the tray base.
8 Turn ON the sampler, close the front door.
9 The system will reset automatically.
10 In the user interface perform a “Sample Transport Self Alignment”.