EasyManuals Logo

Agilent Technologies G1377A User Manual

Agilent Technologies G1377A
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #275 background imageLoading...
Page #275 background image
Specifications 9
1100 Series WS MWS Reference Manual 275
Tabl e 80 Performance Specifications Agilent 1100 Series Micro Well Plate Sampler
Type Specification
GLP features Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications Controller-area network (CAN). RS232C, APG-remote standard,
optional four external contact closures and BCD vial number output
Safety features Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
Injection range 0.01 8 µl in 0.01 µl increments with the small loop capillary
0.01 40 µl in 0.01 µl increments with the extended loop capillary
Precision Typically < 0.5 % RSD of peak areas from 5 40 µl,
Typically <1%RSD from 1–5µl
Typically <3%RSD from 0.2–1µl
Sample viscosity range 0.2 5 cp
Sample capacity 2 × well-plates (MTP) + 10 × 2 ml vials
100 x 2 ml in one tray
40 x 2 ml in half tray
Injection cycle time Typically < 30 s using following standard conditions:
Default draw speed: 4 µl/min
Default eject speed: 10 µl/min
Injection volume: 0.1 µl
Carry-over Typically < 0.05 % using the following conditions:
Column: 150 x 0.5 mm Hypersil ODS, 3 µm
Mobile phase: Water/Acetonitrile = 85/15
Column Flow rate: 13 µl/min
Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test
carryover
Outside wash of needle before injection: 20 sec with water using
flush port

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1377A and is the answer not in the manual?

Agilent Technologies G1377A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1377A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals