270 1100 Series WS MWS Reference Manual
8 Control Module Screens of the Well-plate Samplers
Use the Esc-key to receive Views on the F5-key. Choose System from the
pull-down menu. Use the F3-key (Records) to select the WPS. Use the F5-key
(FW Update) to enter the Update section. Select the a file from the PCMCIA
card (1367nnnn.DLB for the non-thermostatted WPS or 1368nnnn.DLB for the
thermostatted WPS, use 1377nnnn.DLB for the non-thermostatted micro WPS
or 1378nnnn.DLB for the thermostatted micro WPS) and press execute. When
the update has finished the update, press F8-key (Transfer) to return the
module into the normal mode (LED on module should stays yellow).
Use the F5-key (FW Update) to enter the Update section. Select the a file from
the PCMCIA card (1367nnnn.DLB for the non-thermostatted WPS or
1368nnnn.DLB for the thermostatted WPS, use 1377nnnn.DLB for the
non-thermostatted micro WPS or 1378nnnn.DLB for the thermostatted micro
WPS) and press execute. When the update has finished, press F7-key
(Transfer) to return the module into the normal mode (LED on module should
stay yellow).