Agilent 1290 Infinity Binary Pump User Manual 243
LAN Configuration
13
Automatic configuration with Bootp
Adding each instrument to the network using BootP
1 Follow Start > All Programs > Agilent BootP Service and select Edit BootP
Settings. The BootP Settings screen appears.
2 Uncheck the Do you want to log BootP requests? once all instruments have been
added.
The Do you want to log BootP requests? box must be unchecked when you have
finished configuring instruments; otherwise, the log file will quickly fill up
disk space.
3 Click Edit BootP Addresses... The Edit BootP Addresses screen appears.
4 Click Add... The Add BootP Entry screen appears.
Figure 45 Enable BootP logging
5 Make these entries for the instrument:
MAC address
Host name, Enter a Hostname of your choice.
The Host Name must begin with "alpha" characters (i.e. LC1260)
IP address
Comment (optional)