244 Agilent 1290 Infinity Binary Pump User Manual
13 LAN Configuration
Automatic configuration with Bootp
Subnet mask
Gateway address (optional)
The configuration information entered is saved in the Tab File.
6 Click OK.
7 Leave Edit BootP Addresses by pressing Close.
8 Exit BootP Settings by pressing OK.
9 After each modification of the BootP settings (i.e. EditBootPSettings) a stop
or start of the BootP service is required for the BootP service to accept
changes. See “Stopping the Agilent BootP Service” on page 245 or
“Restarting the Agilent BootP Service” on page 246.
10 Power cycle the Instrument.
OR
If you changed the IP address, power cycle the instrument for the changes
to take effect.
11 Use the PING utility to verify connectivity by opening a command window
and typing:
Ping 192.168.254.11 for example.
The Tab File is located at
C:\Documents and Settings\All Users\Application Data\Agilent\BootP\TabFile

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity Binary Pump VL 1290 and is the answer not in the manual?

Agilent Technologies Infinity Binary Pump VL 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity Binary Pump VL 1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals