38 Agilent 1290 Infinity Binary Pump User Manual
3 Installing the Module
Flow connections to the pump
6 Label the tubes accordingly using the supplied stickers and fix the tubes in
the clips of solvent cabinet and 1290 Infinity Binary Pump.
7 Connect the outlet of the Jet Weaver to the autosampler.
8 Connect Waste tubes (G4220-67000) to the purge valve outlets at ports 5
and 6.
9 Connect the corrugated waste tube to the outlet of the leak panel.
10 Route the corrugated waste tube to a waste container.
11 Route drain tubes coming from modules on top of the pump through the
pump.
12 Purge your system prior to the first use (see “Priming the Pump” on
page 54).