50 34410A/11A/L4411A User’s Guide
2Features and Functions
Remote Interface Operation: The Front/Rear switch is manually switchable only,
and cannot be controlled from the remote interface. The following query
returns the current switch setting:
ROUTe:TERMinals?
The query returns either “FRON” or “REAR
Multimeter Reset
The reset function resets most multimeter settings to their factory settings,
except if the power–on state has been manually set to a stored state (see
“Multimeter State Storage” on page 48). The reset function is faster than
cycling power because the power–on self tests are omitted.
Certain settings such as I/O addresses are stored in non–volatile memory,
and are not affected by a reset.
Refer to Power-On and Reset State” on page 85 for a complete listing of the
settings affected by a (front panel or remote interface) reset command.
Front Panel Operation: To reset the multimeter, press . RESET
DMM is displayed, along with the options NO (default) and YES. Select YES to
reset the multimeter.
Remote Interface Operation: The following commands reset the multimeter to
the instrument reset state:
SYSTem:PRESet
*RST
Refer to the Agilent 34410A/11A/L4411A Programmer’s Reference Help for
a complete description and syntax for these commands.
UG_ED5.book Page 50 Thursday, March 1, 2012 11:28 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies L34410A and is the answer not in the manual?

Agilent Technologies L34410A Specifications

General IconGeneral
BrandAgilent Technologies
ModelL34410A
CategoryMultimeter
LanguageEnglish

Related product manuals