108 Chapter 3 
Mode
Occupied Bandwidth Measurement
Mode
Avg Number 
Sets the average count number (the number of sweeps) used with measurement averaging. 
Avg Type 
Accesses averaging settings that determine how the signal will be measured. You can set 
the measurement to logarithmic (video) scaling, or power (RMS) scaling.
Log-Pwr Avg (Video)  Sets the logarithmic (decibel) scale for all filtering and averaging 
processes. This scale is also referred to as “Video” because it is the most common display 
and analysis scale for the video signal within a spectrum analyzer.
Mode SA
Remote Command [:SENSe]:OBWidth:AVERage:COUNt <integer>
[:SENSe]:OBWidth:AVERage:COUNt?
Example AVER:OBW:COUN 50
Preset 100
Min 1
Max 4096
Key Path
Meas Setup, Avg Number 
Mode SA
Remote Command [:SENSe]:OBWidth:AVERage:TYPE RMS|LOG
[:SENSe]:OBWidth:AVERage:TYPE?
Example AVER:OBW:TYPE RMS
Preset LOG
Preset Log-Pwr Avg (Video)
State Saved Saved in instrument state.
Range Log-Pwr Avg (Video) | Pwr Avg (RMS)
Key Path
Meas Setup, Avg Type 
Example AVER:OBW:TYPE LOG