148 Chapter 3 
Mode
Adjacent Channel Power—ACP (I&M) Measurement
Mode
Frequency Low Limit  Sets the negative frequency error limit for the adjacent channel power 
measurement.
Trace/Detector
Accesses the trace and detector settings that can be changed when using the ACP 
measurement.
Detector Avg/Peak
Sets the detector function to average or peak detector. For pulsed signals, you can get a 
more accurate measurement of the signal’s peak or average power by using the Peak 
Power Detector option in the analyzer. 
The analyzer measures average power of signals by default. However, for 
non-constant-duty-cycle signals such as GSM, an average power reading fails to accurately 
show amplitude when the signal is on and if other timeslots in the frame are powered off. 
If the input signal is not bursted, it is probably better to select Average for a more accurate 
result.
State Saved Saved in State
Min 0 Hz
Max Maximum Span
Mode Channel Analyzer
Preset –100 kHz
State Saved Saved in State
Min Maximum Span
Max 0 Hz
Remote Command [:SENSe]:ACPower:DETector[:FUNCtion] AVERage|POSitive
[:SENSe]:ACPower:DETector[:FUNCtion]?
Preset AVERage
State Saved Saved in instrument state.
Range Avg | Pk
Key Path
Trace/Detector, Detector Avg/Peak