38 Chapter 3
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Demod
There is no functionality for this key in this release.
FREQ Channel
Displays the menu of frequency functions.
Remote Command [:SENSe]:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
[:SENSe]:BANDwidth|BWIDth[:RESolution]AUTO
OFF|ON|0|1
[:SENSe]:BANDwidth|BWIDth[:RESolution]AUTO?
Example BAND:RES 3MHz
BAND:RES:AUTO OFF
Restrictions and Notes If an unavailable bandwidth is entered with the numeric
keypad, the closest available bandwidth is selected.
Dependencies /
Couplings
Sweep time in non-zero span is coupled to RBW. As the RBW
changes, the sweep time is changed to maintain amplitude
calibration.
When Res BW is set to Auto, resolution bandwidth is
autocoupled to span. The factory default for the Span to RBW
ratio is approximately 106:1 when auto coupled. When Res
BW is set to Man, bandwidths are entered by the user, and
these bandwidths are used regardless of other analyzer
settings.
Preset 5 MHz, Auto
State Saved Saved in instrument state.
Range 10 Hz to 200 kHz, plus bandwidths of 250 kHz, 300 kHz,
1 MHz, 3 MHz, or 5 MHz
Key Path
BW, Res BW
Mode: N/A
Key Path Front-panel key

Table of Contents