62 Chapter 3
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Segment (Center, 1st or 2nd)
Selects the
Center, 1st or, 2nd segment for limit mask 1 or 2. With each segment setting, the Limits
menu is updated to show the span and level settings available for the currently selected segment
Center Segment
Sets the state of the center segment to On or Off. This key is available only when Center has been
selected with the Segments key.
Range 1 | 2 | 3 | 4
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Limit Trace
Mode SA
Dependencies /
Couplings
Key menu structure is dependant on selected segment.
Remote Command
Notes
The Limit Mask SCPI command syntax uses:
The keyword “CENTer” to define the Center segment level and
span.
The keyword “FIRSt” to define the 1
st
segments left and right
level and span.
The keyword “SECond” to define the 2
nd
segments left and right
level and span.
Preset Center
State Saved Saved in instrument state.
Range Center | 1st | 2nd
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Segment (Center, 1st or 2nd)
Mode SA
Remote Command :CALCulate:LMASk[1]|2:SEGMent:CENTer[:STATe]
ON|OFF|1|0
:CALCulate:LMASk[1]|2:SEGMent:CENTer[:STATe]?
Example :CALC:LMAS1:SEGM:CENT ON
Restrictions and Notes Key is only displayed when the selected segment is Center.