136 Chapter 3
Mode
Adjacent Channel Power—ACP (I&M) Measurement
Mode
Off Turns averaging functions off.
Exponential Sets the averaging function to exponential. With exponential averaging mode,
each successive data acquisition is exponentially weighted and combined with the existing
average. This weights new data more than the old data, which facilitates tracking of slow
changing signals.
Max Hold In
Max Hold the analyzer maintains and displays a max hold trace, which
represents the maximum data value on a point-by-point basis of the new trace data and
previous trace data. Pressing the
Max Hold key for the selected trace sets the trace type to
Max Hold, causes the trace to be cleared and causes the Max Hold sequence to be restarted.
Preset The averaging state default is OFF. If averaging was turned
ON, the default averaging mode for the measurement would be
EXPonential.
State Saved Saved in State
Range Exponential | Max Hold | Repeat | RMaxHold
Key Path
Meas Setup, Avg Mode
Mode Channel Analyzer
Remote Command [:SENSe]:ACPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:AVERage[:STATe]?
Example ACP:AVER OFF
Dependencies /
Couplings
If Averaging State is turned ON, the Averaging Mode used per
default by the measurement is Exponential.
Preset OFF
State Saved Saved in State
Key Path
Meas Setup, Avg Mode, Off
Example ACP:AVER:TCON EXP
Key Path
Meas setup, Avg Mode, Exponential
Example ACP:AVER:TCON MAXH