136 Chapter 3 
Mode
Adjacent Channel Power—ACP (I&M) Measurement
Mode
Off  Turns averaging functions off.
Exponential  Sets the averaging function to exponential. With exponential averaging mode, 
each successive data acquisition is exponentially weighted and combined with the existing 
average. This weights new data more than the old data, which facilitates tracking of slow 
changing signals.
Max Hold  In 
Max Hold the analyzer maintains and displays a max hold trace, which 
represents the maximum data value on a point-by-point basis of the new trace data and 
previous trace data. Pressing the 
Max Hold key for the selected trace sets the trace type to 
Max Hold, causes the trace to be cleared and causes the Max Hold sequence to be restarted.
Preset The averaging state default is OFF. If averaging was turned 
ON, the default averaging mode for the measurement would be 
EXPonential.
State Saved Saved in State
Range Exponential | Max Hold | Repeat | RMaxHold
Key Path
Meas Setup, Avg Mode 
Mode Channel Analyzer
Remote Command [:SENSe]:ACPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:AVERage[:STATe]?
Example ACP:AVER OFF
Dependencies / 
Couplings
If Averaging State is turned ON, the Averaging Mode used per 
default by the measurement is Exponential.
Preset OFF
State Saved Saved in State
Key Path
Meas Setup, Avg Mode, Off
Example ACP:AVER:TCON EXP
Key Path
Meas setup, Avg Mode, Exponential
Example ACP:AVER:TCON MAXH