Chapter 3 131
Mode
Adjacent Channel Power—ACP (I&M) Measurement
Mode
Control/Sweep
Accesses the sweep and measurement control settings that can be changed when using the 
ACP measurement.
Restart, Single, Cont
The sweep control functionality is shared across all the Modes. See “Control/Sweep 
Functions” on page  252 for details about how they operate. 
Sweep Time Query (Remote Command Only)
The sweep time value can be queried using the [:SENSe]:ACPower:SWEep:TIME? 
command. The ACP measurement is a frequency domain measurement only, so the Sweep 
Time key is disabled (not available) in the ACP measurement. 
Example :DISP:ACP:WIND:TRACe:Y:PDIV 10dB
Preset 10.00 dB / Div
State Saved Saved in State
Range 1.0 to 20
Key Path
AMPTD Y Scale, Scale/Div 
Mode: SA
Key Path Front-panel key
Mode SA
Remote Command [:SENSe]:ACPower:SWEep:TIME?
Example ACP:SWE:TIME?
Notes In ACP measurement (Frequency Domain measurement), you 
can only query the “sweep” time. Trying to set the Sweep Time 
will generate a system error message
Preset The default value is auto-calculated and will depend on the 
Frequency range of the Analyzer.
Key Path N/A - Remote Feature Only