Chapter 3 119
Mode
Occupied Bandwidth Measurement
Mode
The default setting is High, which optimizes dynamic range but may introduce distortion 
with large signal levels. Setting sensitivity to Low removes signal distortion, but may 
decrease dynamic range.
Trigger
The Trigger features are only available in zero span. In non-zero span, the trigger type is 
always free run.
Power 
Sets the percentage of total input signal power used when determining the occupied 
bandwidth.
Peak Search
Accesses the marker peak functions and places a marker on the highest peak of the 
currently selected trace.
Mode SA
Remote Command [:SENSe]:OBWidth:POWer:SENSitivity LOW|HIGH
[:SENSe]:OBWidth:POWer:SENSitivity?
Preset High
Range Low | High
Key Path
Meas Setup, Sensitivity 
Mode SA
Remote Command [:SENSe]:OBWidth:PERCent <percent>
[:SENSe]:OBWidth:PERCent?
Preset 99%
State Saved Saved in instrument state.
Min 10
Max 99.9
Key Path
Meas Setup, Power