Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Auto Range 
For the “Electronic Attenuation” parameter, when the “Auto Range” is ON the entire 
frequency range of the analyzer is scanned to determine the total power applied to the 
analyzer. The RF input attenuation is then automatically set to ensure that the analyzer 
input is not overloaded.
For the “Internal Preamplifier” parameter, turning on the preamp gives a better noise 
figure, but a decreased TOI to noise floor dynamic range. So when the “Auto Range” is ON, 
the instrument will optimize this setting.
Y Axis Units 
Displays the menu keys that enable you to change the vertical (Y) axis amplitude units. 
The analyzer retains the entered Y Axis Units separately for both Log and Lin amplitude 
scale types. For example, you can set 
Scale Type to Log and Y Axis Units to dBmV. Then you 
can set 
Scale Type to Lin and Y Axis Units to W. When you switch between Log and Lin the 
Y Axis Units will remain to dBmV in Log and W in Lin. 
When you preset the analyzer, Scale Type is set to Log with dBm default Y Axis units. 
When you switch the Scale Type to Lin, the Y Axis units default to Volts.
Range Log | Lin
Key Path
AMPTD Y Scale, Scale Type 
Mode SA
Remote Command [:SENSe]:POWer[:RF]:ARANge ON|OFF|1|0
[:SENSe]:POWer[:RF]:ARANge?
Example POW:ARAN OFF
Preset Off
Range On | Off | 1 | 0
Key Path
AMPTD Y Scale, Auto Range 
Mode SA
Remote Command :UNIT:POWer DBM|DBMV|DBUV|V|W|A
:UNIT:POWer?
Preset dBm (
Scale Type = Log) 
Range dBm | dBmV | dBuV | Watts | Volts | Amps