70 Chapter 3 
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Level: 2nd Segment Right  
Sets the power level, for the 2
nd
 segment to the Right of the Center segment. This key is only 
available when 
2
nd
 has been selected with the Segments key.
Sensitivity 
The sensitivity setting changes how the auto-range algorithm is applied to input signals. 
The default setting is High, which optimizes dynamic range but may introduce distortion 
with large signal levels. Setting sensitivity to Low removes signal distortion, but may 
decrease dynamic range.
Default Unit dB
Mode SA
Remote Command :CALCulate:LMASk[1]|2:SECond:SEGMent:RIGHt:LEVel 
<ampl>
:CALCulate:LMASk[1]|2:SECond:SEGMent:RIGHt:LEVel?
Example :CALC:LMAS:SEGM:SEC:RIGH:LEV –39 dBm
Notes
Key is only displayed when the selected segment is 2
nd
, and the 
selected segment side is Right.
Dependencies / 
Couplings
Value is displayed on the key in the currently selected Y Axis 
Unit.
Preset Not affected by a preset but set per default to –20 dBm after a 
power cycling.
State Saved Saved in instrument state.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Right, Level
Default Unit dBm
Mode SA
Remote Command [:SENSe]:POWer:SENSitivity LOW|HIGH
[:SENSe]:POWer:SENSitivity?
Preset High
Range Low | High
Key Path
Meas Setup, Sensitivity