76 Chapter 3 
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Span/X Scale (Spectrum Analyzer Measurement)
Activates the span functions menu. The menu selections are determined by the current 
measurement mode. 
Span 
Sets the span frequency value. Span frequency describes the frequency range that is 
positioned symmetrically about center frequency.
Changing span changes the displayed frequency range symmetrically about the center 
frequency. 
While in the SA measurement, setting the span to 0 Hz through SCPI or the front-panel 
numeric key pad puts the analyzer into the Zero Span Measurement. 
While in the Zero Span measurement, setting the Span to a non-zero value through SCPI 
or Front Panel puts the analyzer in the SA measurement.
In a Zero Span measurement, the frequency entry mode is always center/span.
Mode All
Preset Full span (depends on hardware present – options)
State Saved Saved in State
Key Path Front-panel key
Mode SA
Remote Command [:SENSe]:FREQuency:SPAN <frequency>
[:SENSe]:FREQuency:SPAN?
Example FREQ:SPAN 2 GHz sets the span to 2 GHz
FREQ:SPAN 0 Hz sets the span to 0 Hz and puts the 
instrument in Zero Span measurement mode.
Dependencies / 
Couplings
Span affects RBW
Remote Command 
Notes
Preset and Max values are dependent on Hardware (Options)
Preset 2.9999 GHz (Option 503)
5.9999 GHz (Option 506)
State Saved Saved in State