Chapter 3 57
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Marker ->
Accesses the marker function menu keys that change value of center frequency or 
reference level to the current value of the selected marker.
Mkr -> CF
Sets the center frequency equal to the specified marker frequency, which moves the 
marker to the center of the screen. In delta marker mode, the center frequency is set to the 
frequency value of the delta marker. This feature is not available in zero span.
Mkr -> RL
Sets the reference level to the amplitude of the specified marker. In delta marker mode, 
the reference level is set to the amplitude difference between the two markers.
Meas Setup
In the Spectrum Analysis mode, Meas Setup accesses a menu of keys associated with the 
current measurement. the following keys for setting up a signal measurement, press 
Meas, 
Meas Off and then the Meas Setup key.
Mode Spectrum Analyzer
Key Path
Marker ->
Mode Spectrum Analyzer
Remote Command :CALCulate:MARKer[1]|2|3|4[:SET]:CENTer
Example CALC:MARK2:SET:CENT
Key Path
Marker ->, Mkr -> CF
Mode Spectrum Analyzer, Stimulus Response
Remote Command :CALCulate:MARKer[1]|2|3|4[:SET]:RLEVel
Example CALC:MARK2:SET:RLEV 
Key Path
Marker ->, Mkr -> RL