118 Chapter 3
Mode
Occupied Bandwidth Measurement
Mode
Sets the power level, for the 2
nd
segment to the Left of the Center segment. This key is only
available when
2
nd
has been selected with the Segments key.
Level: 2nd Segment Right
Sets the power level, for the 2
nd
segment to the Right of the Center segment. This key is only
available when
2
nd
has been selected with the Segments key.
Sensitivity
The sensitivity setting changes how the auto-range algorithm is applied to input signals.
Mode Spectrum Analyzer
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Left.
Preset Not affected by a preset but set per default to 10 dBm after a
power cycling
State Saved Saved in instrument state.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Left Level
Default Unit dB
Mode SA
Remote Command :CALCulate:OBWidth:LMASk[1]|2:SECond:SEGMent:RIGHt:L
EVel <ampl>
:CALCulate:OBWidth:LMASk[1]|2:SECond:SEGMent:RIGHt:L
EVel?
Example :CALC:OBW:LMAS:SEGM:SEC:RIGH:LEV –39 dBm
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Right.
Dependencies /
Couplings
Value is displayed on the key in the currently selected Y Axis
Unit.
Preset Not affected by a preset but set per default to –10 dBm after a
power cycling.
State Saved Saved in instrument state.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Right Level
Default Unit dBm