Chapter 3 95
Mode
Occupied Bandwidth Measurement
Mode
selected, the vertical divisions are scaled in logarithmic units. Scale Type (Lin) scales the
vertical divisions linearly, and sets the default amplitude units to volts.
Y Axis Units
Displays the menu keys that enable you to change the vertical (Y) axis amplitude unit. The
analyzer retains the entered Y Axis Units separately for both Log and Lin amplitude scale
types. For example, if
Scale Type has been set to Log, and you set Y Axis Units to dBm,
pressing
Scale Type (log) sets the Y Axis Units to dBm. If Scale Type has been set to Lin and
you set
Y Axis Units to V, pressing Scale Type (Lin) sets the Y Axis Units to V. Pressing Scale
Type (Log)
again sets the Y Axis Units back to dBm.
dBm Sets the amplitude unit for the selected amplitude scale (log/lin) to dBm.
Mode SA
Remote Command :DISPlay:OBWidth:WINDow:TRACe:Y[:SCALe]:SPACing
LINear|LOGarithmic
:DISPlay:OBWidth:WINDow:TRACe:Y[:SCALe]:SPACing?
Example DISP:OBW:WIND:TRAC:Y:PDIV LIN
Preset Log
State Saved Saved in State
Range Log | Lin
Key Path
AMPTD Y Scale, Scale Type
Mode SA
Remote Command :UNIT:OBWidth:POWer DBM|DBMV|DBUV|V|W|A
:UNIT:OBWidth:POWer?
Example UNIT:OBW:POW DBMV
Preset dBm (
Scale Type = Log)
Range dBm | dBmV | dBuV | Watts | Volts | Amps
Key Path
AMPTD Y Scale, Y Axis Units
Example UNIT:OBW:POW DBM
Key Path
AMPTD Y Scale, Y Axis Units, dBm