116 Chapter 3
Mode
Occupied Bandwidth Measurement
Mode
2nd Segment
Sets the state of the first segment to On or Off. This key is available only when 2
nd
has been
selected with the Segments key.
Remote Command :CALCulate:OBWidth:LMASk[1]|2:SEGMent:FIRSt:RIGHt:SP
AN <frequency>
:CALCulate:OBWidth:LMASk[1]|2:SEGMent:FIRSt:RIGHt:SP
AN?
Example :CALC:OBW:LMAS1:SEGM:FIRS:RIGH:SPAN 1.75 MHz
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Right.
Dependencies /
Couplings
If the Center, 1
st
, and 2
nd
segments On, the right 2
nd
segment
span adjusts on a change in the right 1
st
segment span.
Preset Not affected by a preset but set per default to 2 MHz after a
power cycling
State Saved Saved in instrument state.
Min 0 Hz
Max See Dependencies / Couplings “1st Segment” on page 112.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Right Span
Default Unit Hz
Mode SA
Remote Command :CALCulate:OBWidth:LMASk[1]|2:SEGMent:SECond
ON|OFF|1|0
:CALCulate:OBWidth:LMASk[1]|2:SEGMent:SECond?
Example :CALC:OBW:LMAS1:SEGM:SEC ON
Notes
Key is only displayed when the selected segment is 2
nd
.
Dependencies /
Couplings
If Off the 2
nd
limit mask segment will not be displayed even if
turned On.
If On, and the First and Center segments are On:
The left hand side 2
nd
segment is drawn to the left hand side of
the left first segment at the specified power level
The right hand side 1
st
segment is drawn to the right hand side
of the right first segment at the specified power level