Chapter 3 69
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Level: 2nd Segment Left
Sets the power level, for the 2
nd
segment to the Left of the Center segment. This key is only
available when
2
nd
has been selected with the Segments key.
Dependencies /
Couplings
The Left span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment left span>
The Right span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment right span>
Where <maximum span> is the maximum frequency span of
the instrument (3 GHz or 6 GHz)
Remote Command
Notes
The Limit Mask SCPI command syntax uses:
The keyword “LEFT” to define the level and span for the second
segment to the Left of the center segment.
The keyword “RIGHt” to define the level and span for the
second segment to the Right of the center segment.
Preset Not affected by a preset but set per default to Left after a power
cycling.
State Saved Saved in instrument state.
Range Left | Right
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment (Left or Right)
Mode Spectrum Analyzer
Remote Command :CALCulate:LMASk[1]|2:SECond:SEGMent:LEFT:LEVel
<ampl>
:CALCulate:LMASk[1]|2:SECond:SEGMent:LEFT:LEVel?
Example :CALC:LMAS:SEGM:SEC:LEFT:LEV –39 dBm
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Left.
Preset Not affected by a preset but set per default to -20 dB after a
power cycling
State Saved Saved in instrument state.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 2nd Segment: Left, Level