Chapter 3 105
Mode
Occupied Bandwidth Measurement
Mode
Meas
The Meas key displays a menu that lets you make one-button measurements such as the
occupied bandwidth measurement. In the Spectrum Analyzer mode (see “Spectrum
Analyzer” on page 28), Measurements Off is the default measurement. You must select
Occupied BW. The measurement runs using the current default setup. Or you can change
settings for your unique measurement requirements.
Settings can be changed for these measurements using the functions under the Meas
Setup key (see “Meas Setup” on page 105).
Meas Setup
In the Spectrum Analyzer mode, Meas Setup accesses a menu of keys associated with the
current measurement. To access the following keys for setting up an Occupied BW
measurement, you must have that measurement selected. Press
Meas, Occupied BW to
select the measurement.
Avg Mode
Averaging functions are the same as defined in the Spectrum Analyzer measurement, but
the parameter values are measurement dependent. Therefore, the remote commands are
different.
Sets the type of termination control used for the averaging function. This determines the
averaging action after the specified number of measurements (see
Avg Number) has been
reached.
Mode: All
Key Path Front-panel key
Mode: SA
Key Path Front-panel key
Mode SA
Remote Command [:SENSe]:OBWidth:AVERage:TCONtrol
EXPonential|REPeat|RMAXhold
[:SENSe]:OBWidth:AVERage:TCONtrol?
Example AVER:OBW:TCON REP