Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
running – the trace is not updated until the last average count has been
reached.
Repeat Max Hold Sets
Avg Mode to repeat with max hold function. Repeat with max hold
captures and holds the highest magnitude data at the end of each sweep count.
NOTE With large average counts and/or slow traces, the trace from the last
completed result remains on the display while
Repeat Max Hold is
running – the trace is not updated until the last average count has been
reached. (Maximum data points are held with each consecutive sweep)
Avg Number
Sets the average count number (the number of sweeps) used with measurement averaging.
Avg Type
Accesses averaging settings that determine how the signal will be measured. You can set
Example AVER:TCON REP
Key Path
Meas Setup, Avg Mode, Repeat
Example AVER:TCON RMAX
Key Path
Meas Setup, Avg Mode, Repeat Max Hold
Mode SA
Remote Command [:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Example AVER:COUN 50
Preset 100
Min 1
Max 4096
Key Path
Meas Setup, Avg Number