Chapter 3 67
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Span: 1st Segment Right
Sets the frequency span for the 1
st
segment to the Right of the Center segment. This key is only
available when
1st has been selected with the Segments key.
Notes
Key is only displayed when the selected segment is 1
st
, and the
selected segment side is Left.
Dependencies /
Couplings
If the Center, 1
st
, and 2
nd
segments are On, the left 2
nd
segment
span adjusts on a change in the left 1
st
segment span.
Preset Not affected by a preset but set per default to 2 MHz after a
power cycling
State Saved Saved in instrument state.
Min 0 Hz
Max Current measurement span.
See Dependencies / Couplings “1st Segment” on page 64.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Left, Span
Default Unit Hz
Mode SA
Remote Command :CALCulate:LMASk[1]|2:SEGMent:FIRSt:RIGHt:SPAN
<frequency>
:CALCulate:LMASk[1]|2:SEGMent:FIRSt:RIGHt:SPAN?
Example :CALC:LMAS1:SEGM:FIRS:RIGH:SPAN 1.75 MHz
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Right.
Dependencies /
Couplings
If the Center, 1
st
, and 2
nd
segments On, the right 2
nd
segment
span adjusts on a change in the right 1
st
segment span.
Preset Not affected by a preset but set per default to 2 MHz after a
power cycling
State Saved Saved in instrument state.
Min 0 Hz
Max Current measurement span.
See Dependencies / Couplings “1st Segment” on page 64.
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Right, Span
Default Unit Hz